Horizontal Steam Sterilizer is a horizontal autoclave for disinfecting and sterilizing heat-resistant medical and laboratory materials with saturated steam.It supports electric or direct steam heating and integrates chamber, jacket, piping, and control systems.
Technical Specifications
|
Parameter |
WS-150YDB |
|
Chamber Volume |
150 L |
|
Chamber Size |
Ø440 × 1000 mm |
|
Working Pressure |
0.22 MPa |
|
Working Temperature |
134°C |
|
Temperature Range |
105–134°C |
|
Sterilization Time |
0–99 min |
|
Drying Time |
0–99 min |
|
Temperature Uniformity |
≤ ±2°C |
|
Heating Power |
9 kW |
|
Power Supply |
380 V, 50 Hz |
|
Overall Dimensions |
1400 × 650 × 1600 mm |
|
Transport Dimensions |
1560 × 820 × 1820 mm |
|
Net / Gross Weight |
340 / 430 kg |
Product Characteristics
Automatic Cycle Control: The sterilization process is controlled automatically by the computer control system, reducing the need for manual adjustment during a cycle.
Integrated Drying Function: A programmable drying stage can be set for 0–99 minutes and is suitable for drying medical dressings after sterilization.
Built-in Safety Protection: The unit includes over-temperature, over-pressure and low-water protection, with automatic alarm and power cut-off under low-water conditions.
Cycle Printing: An integrated printer is provided for recording sterilization cycle information for operational documentation.
Sterilization System & Cycle Control
Saturated Steam Sterilization: Sterilization is performed using saturated steam, with a working temperature of 134°C and working pressure of 0.22 MPa for the listed WS-150YDB model.
Gravity Air Removal: The chamber uses gravity displacement to release cold air before sterilization, supporting steam contact with the load.
Automatic Steam Regulation: The control system adjusts steam inlet and outlet according to chamber temperature during the sterilization process.
Programmable Time Settings: Sterilization and drying times can each be set from 0–99 minutes, allowing the cycle duration to be adjusted to the load requirement.
Sterilizer Configuration & Structure
Chamber Pressure Monitoring: A dedicated pressure gauge displays the pressure inside the sterilization chamber during operation.
Jacket Pressure Monitoring: A separate pressure gauge monitors jacket pressure, allowing the heating-side pressure to be checked independently.
Door & Handle Assembly: The front door is fitted with an opening handle for loading and unloading sterilized materials.
Control & Display Panel: The control system integrates an LED display, power switch and printer for equipment operation and cycle record output.
Applications & Load Compatibility
Hospital & Clinic Instruments: Suitable for steam sterilization of reusable surgical instruments and other heat-resistant medical equipment.
Laboratory Materials: Applicable to glassware, laboratory utensils and culture media used in scientific research and laboratory operations.
Mushroom & Culture Media: Suitable for sterilizing culture media and glass containers used in mushroom cultivation and related laboratory processes.
Heat-Resistant Fabric Loads: Can be used for steam sterilization and subsequent drying of heat-resistant medical dressings and fabric materials.
Quality Control & Factory Testing
Chamber Inspection: The stainless steel sterilization chamber and main structural components are inspected before final assembly.
Control System Check: The automatic control system is checked for cycle operation, temperature adjustment and alarm response.
Safety Function Check: Door interlock, over-pressure protection and low-water protection are checked before shipment.
Operating Cycle Test: The assembled unit is run through an operating cycle to verify heating, steam control, sterilization timing and drying functions.
Installation & Utility Requirements
Electrical Supply: The WS-150YDB requires a 380 V, 50 Hz power supply with a rated heating power of 9 kW.
Heating Source: The sterilizer supports electric heating or direct steam heating. For direct steam operation, the required steam supply conditions should be confirmed before installation.
Installation Space: The machine measures 1400 × 650 × 1600 mm, while the transport dimensions are 1560 × 820 × 1820 mm; sufficient clearance should be reserved for loading and servicing.
Equipment Handling: With a net weight of approximately 340 kg, suitable lifting and positioning equipment should be arranged for installation and relocation.


Packaging

FAQ
Q: What information is needed for a quotation?
A: Please provide order quantity, destination country, and any specific configuration or documentation requirements.
Q: Can technical documents be provided before ordering?
A: Available product specifications, drawings, manuals, and related documents can be provided for review.
Q: Can spare parts be ordered separately?
A: Yes. Replacement parts can be supplied according to the specific equipment configuration.
Q: Can spare parts be included with the initial order?
A: Required spare parts can be quoted together with the machine before production.
Q: Is private labeling available for distributors?
A: Nameplates and product labeling can be discussed according to distributor requirements.
Q: What is the production lead time?
A: Lead time is based on order quantity, configuration, and current production scheduling.
Q: Is technical support available after delivery?
A: Remote technical assistance can be arranged according to the equipment and service requirements.
Hot Tags: horizontal steam sterilizer, China horizontal steam sterilizer suppliers, factory, autoclave, Horizontal Steam Sterilizer